Note that the following uses pre-cast gels and pre-made running buffer, see accessory protocols [NotDoneYetDudez] for casting gels and making your own running & loading buffers
Supplies:
~1L of 1x Invitrogen MES PAGE Running Buffer (supplied at 20x, Life Tech. Cat. No. XXXX)
1 or 2 precast acrylamide gels (I like the 4-20% gradient gels, e.g. Life Tech Cat. No. XXXX)
For protein solutions (e.g. column flow-through, cleared lysate, purified protein):
Add 1/3 vol. 4x SDS-PAGE Sample Loading Buffer and pipette up and down to mix
Incubate sample at 95degC 5min
I like to do this step in the thermocycler; prepare < 50uL of sample + loading buffer in a 0.2mL PCR tube or strip and incubate at least 5min @ 95C
For E. Coli cultures (e.g. to check protein expression):
Remove a small amount of culture (25uL is good) DON'T spin down
Add 1 vol. of 4x sample buffer directly to the culture
Incubate 5min @ 95degC in the thermocycler
Dense cultures can become viscous and nearly impossible to load after boiling; if this is the case, repeat above steps with 1-3 more volumes of sample buffer, or dilute culture 2-4x with water
spin tubes briefly to remove condensation from cap
Note that you don't need to boil the pre-stained ladder
Setting up the Gel Rig, Loading, and Running:
Prepare 1x running buffer (50mL of 20x in 1L ddH2O is usually enough for mini gels)
Open gel packages
MAKE SURE TO REMOVE TAPE STRIP FROM BOTTOM OF GEL OR IT WON'T RUN
If running 2 gels, put one on either side of the white electrode assembly, with the short plate facing inwards (gels should be 'facing' one another)
If only running one gel, use the plastic buffer dam in place of the 2nd gel (writing faces inwards)
Remove plastic combs from the tops of the gels
Pour running buffer into top chamber (between gels) until buffer level is higher than the top of the (shorter) inner gel plate (i.e. until buffer covers the wells). Leave for a minute or two to make sure there are no leaks (buffer level doesn't drop).
Fill the outer chamber until ~3/4 of the gel is covered.
[Optional, but recommended] Use a 1mL pipet to wash some running buffer into wells to clear out any gel fragments or unpolymerized acrylamide
Load samples into wells with a P20 tip or gel-loading tip
5uL is usually sufficient for a 1mm gel; I wouldn't load more than 10uL unless you know you have a very dilute sample
Load 2-5uL of Prestained Ladder
Attach lid and run gel!
1hr at 150V is usually sufficient
If running for a western blot, run until dye front has passed completely out of the gel