Microplate Reader Quick Start
Best Practices
- Follow typical best practice for experimental design. Utilize technical replicates from the same biological sample. Have biological replicates spread across different days, appropriate controls, etc.
- It's always best to start from single colony picks that had been grown overnight in liquid culture to saturation.
- A typical start culture dilution is 2:200, 1:200 or 1:250 from an overnight culture. The most diluted one (1:250) would provide the best result for measuring lag phase times.
- It is advisable to spread out samples across the plate, rather than clustering samples together or along the same column/row. This reduces the impact of biases due to evaporation, temperature, or other affects across the plate. Randomization is close to (but not always entirely) ideal. See this paper – the concepts broadly apply.
- Edge wells will experience more evaporation during long experiments incubating at high temperatures. You may want to fill these with media/buffer and not use them in your experiment (not even as blanks). Filling them will "protect" the inner wells to some extent.
- Remember to record the positions of each sample for later analysis and to include at least one blank.
- Select the appropriate type of plate for your application: black walled, clear bottom for fluorescence; white walled, clear bottom for luminescence; and completely clear for other applications, like measuring OD600;
Plate reader operation
- If your plate reader is controlled by a Windows computer (like ours), turn off WiFi and ensure that the computer is up to date prior to starting your experiment. Windows Update will reboot the computer mid-experiment and kill your run.
- Ensure the computer is on and the plate reader is plugged in prior to powering on the plate reader.
- Depending on the software you use (ex. iControl), using the scroll wheel may change a numerical setting. Be cognizant of this and take care not to inadvertently adjust settings
Setting Instrument GAIN for Fluorescence/Luminescence (AUTOMATIC versus MANUAL)

Improperly setting the gain to AUTOMATIC for
time course data is one of the most common mistakes new researchers make!
GAIN is the sensitivity of the photodetector in the plate reader to the emission of light with whatever you are measuring. It is a multiplier between the number of detected photons and the number output by the plate reader as signal. Increasing the GAIN will improve the dynamic range of your measurements and give you data for dim samples; however, too high of a GAIN will result in saturation of the detector making certain values unusable. This is the same problem that happens if you overexpose a photo of a gel so the bands are saturated.
Single time point assays: If you are performing an assay that only needs to make a
measurement at one time point, you can set the GAIN on the plate reader to AUTOMATIC. In this setting, the instrument will automatically test different GAIN values (multipliers) and pick one that makes your brightest samples near the high end of its measurement range (and therefore, all of your values as high as possible without going over the saturation point). The output file will show the value for the GAIN it picked.
Time-course assays. For these experiments, it is very important to use MANUAL mode to set the GAIN yourself. If you set the gain to AUTOMATIC, the instrument will automatically adjust the gain
differently for each time point, rendering your data so that it cannot be compared across time. How do you know what MANUAL setting to use for the GAIN for a growth curve of a fluorescent strain? Take some of the saturated preconditioned sample and put it into the same plate i the same volume as you will use for the time-course assay and run it with AUTOMATIC gain or adjust the gain yourself. Then, it's best to back the gain off a bit: set it to 80% of what was picked by the automatic (for example) so that if your actual measurements are a bit higher, they won't saturate.
Though GAIN can be thought of as a photodetector multiplier, you should not assume that it is linear. This means that you should never measure a plate at one GAIN and then measure it at double the GAIN and assume that your fluorescence values will also double. The GAIN to use can also differ from experiment to experiment depending on the plate reader, the plate used, the measurement in question, media used, and other factors. It's always best to calibrate it in a preliminary experiment versus risk the wrong setting data from a large experiment interpretable!
Standard wavelengths
Different fluorescent markers have different absorption and emission spectra. While there are resources that detail the spectra, such as FPbase, the most appropriate wavelengths to use for measurements may not be immediately apparent. Provided here is a list of commonly used fluorphores and appropriate wavelengths to use.

One common error is to assume that you should use the wavelengths that are the peaks of the excitation and emission spectra. These wavelengths may be too close together for the filters on a machine, resulting in scattering of the excitation light being read as fluorescence emission. For example, do NOT use excitation 488 nm and emission 509 nm for GFP. Numbers in parentheses on an instrument after a wavelength show the total size of the window of wavelengths transmitted by a filter (e.g., 450(20) nm means light gets through the filter 440 to 460 nm).
| Measurement |
Absorption |
Emission |
Citation |
Notes |
| Cell Density |
600 |
600 |
Myers 2013 |
Use 660 with mRFP |
| Cell Density w/ mRFP |
660 |
660 |
Pedelacq 2006 |
Use 660 with mRFP |
| Fluorescent Protein |
Excitation |
Emission |
Citation |
Notes |
| BFP (mTagBFP) |
405 nm |
453 nm |
Radde 2024 |
|
| YFP |
436 nm |
480 nm |
Hegreness 2006 |
|
| GFP (sfGFP) |
485 nm |
528 nm |
Radde 2024 |
|
| CFP |
500 nm |
535 nm |
Hegreness 2006 |
|
| RFP (mRFP1) |
585 nm |
610 nm |
Radde 2024 |
Be aware of absorption overlap with 600 nm for OD measurements |
Our Plate Reader Protocols